Your Agtron number might be lying to you.

Your Agtron number might be lying to you.

Same coffee. Same roast. Different grind → potentially different Agtron reading.

Why? Grind size changes the surface being measured, and therefore how light is reflected back to the instrument.

So before saying, “This roast is darker, different, too light” ask:

Did the roast change, or did the measurement change?

Standardize your grind, sample preparation, and measurement conditions.

Otherwise, you may be measuring grind size, not roast color.

And the difference can be significant. Fine vs. coarse grinds can potentially shift the reading by 20 Agtron points or more.

My QC tips:

• Keep your grind consistent, e.g. ~600–800 µm as a controlled target, rather than changing grind between samples.

• Measure at the same post-roast time.

• Use the same grinder and setting.

• Take 3 measurements from 3 separate samples and look at the average.

• Use Agtron primarily as a color/QC measurement, NOT a substitute for sensory evaluation.

• I would not recommend comparing measurement from different devices, colorimeters may use different optical wavelengths, geometries, and calculation methods

Standardize the measurement before you trust the number.